Using The Built-In Test Chromatogram; Procedure Using The Agilent Labadvisor - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Using the Built-in Test Chromatogram

This function is available from the Agilent ChemStation, LabAdvisor and
Instant Pilot.
The built-in Test Chromatogram can be used to check the signal path from the
detector to the data system and the data analysis or via the analog output to
the integrator or data system. The chromatogram is continuously repeated
until a stop is executed either by means of a stop time or manually.
The peak height is always the same but the area and the retention time depend on the set
N O T E
peakwidth, see examples below.

Procedure Using the Agilent LabAdvisor

This procedure works for all Agilent 1200 Infinity detectors (DAD, MWD, VWD,
FLD and RID). The example figure is from the RID detector.
1 Assure that the default LC method is loaded via the control software.
2 Start the Agilent LabAdvisor software (B.01.03 SP4 or later) and open the
3 Open the test chromatogram screen
4 Turn the Test Chromatogram on.
5 Change to the detector's Module Service Center and add the detector signal to
Agilent 1260 Infinity DAD and MWD User Manual
detector's Tools selection.
the Signal Plot window.

Using the Built-in Test Chromatogram

9
Test Functions
181