Filter Test Evaluation
Filter Test Failed
Test Failed
Probable cause
1
2
Holmium Oxide Maximum out of Limits
Probable cause
1
2
Agilent 1260 Infinity DAD and MWD User Manual
Filter assembly (lever and filter) not installed. Install the filter assembly.
Defective filter motor.
Holmium oxide filter not installed.
Dirty or contaminated filter.
Suggested actions
Exchange the filter motor.
Suggested actions
Install the holmium oxide filter.
Exchange the holmium oxide filter.
9
Test Functions
Filter Test
165