Agilent Technologies 1260 User Manual page 213

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

13
Fix the heat exchanger to the clamp unit and the flow cell
body to the heat exchanger..
15
Check for a centered holder vs. hole. If required adjust
with the holder screws.
Agilent 1260 Infinity DAD and MWD User Manual
Replacing Capillaries on a Standard Flow Cell
14
Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.
Next Steps:
16
Reconnect the capillaries, see
Cell"
on page 200.
17
Perform a leak test.
18
Insert the flow cell.
19
Replace the front cover.
20
Perform a
"Wavelength Verification and Recalibration"
page 183 or a
"Holmium Oxide Test"
the correct positioning of the flow cell.
10
Maintenance
"Exchanging a Flow
on
on page 174 to check
213