Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
Loss of data
C A U T I O N
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnosis and repair log books which will not be erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 11
Forced Cold Start Settings (without on-board LAN)
Mode Select
LAN
TEST/BOOT
No LAN
TEST/BOOT
Agilent 1260 Infinity DAD and MWD User Manual
Save your methods and data before executing a forced cold start.
SW1
SW2
1
1
Setting the 8-bit Configuration Switch
SW3
SW4
1
0
0
1
0
0
Introduction
SW5
SW6
SW7
0
0
0
1
0
0
1
SW8
1
1
31