Cell Test Evaluation - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

9
Test Functions
Cell Test

Cell Test Evaluation

Cell Test Evaluation
The Agilent ChemStation calculates the intensity ratio automatically. The
intensity ratio (typically between 0.5 and 0.7 for new standard flow cells and
0.1 to 0.3 for new mico- and high pressure cells) is dependent on the degree of
contamination of the flow cell windows, and on the type of flow cell used.
Figure 69
This test can be used for the standard flow cells only. The nano flow cells will give very low
N O T E
values due to their design.
Test Failed (low ratio value)
Probable cause
1
2
180
Cell Test Results (report) with no flow cell inserted
Absorbing solvent or air bubble in flow cell.
Dirty or contaminated flow cell.
Suggested actions
Ensure the flow cell is filled with water, and free
from air bubbles.
Exchange the flow cell windows.
Agilent 1260 Infinity DAD and MWD User Manual