Conditions
A single injection of the isocratic test standard is made under the conditions
given in
Table 22
Flow
Stoptime
Solvent
Temperature
Wavelength
Injection Volume
Agilent 1260 Infinity DAD and MWD User Manual
Table 22
on page 87:
Conditions
1.5 ml/minute
8 minutes
100% (30% water/70% Acetonitrile)
Ambient
sample 254 nm (4 nm bandwidth) reference 360 nm (100 nm bandwidth)
1 µl