Agilent Technologies 1260 User Manual page 222

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

10
Maintenance
Nano Flow Cell - Replacing or Cleaning
The cell body can be fitted in two positions to allow the capillaries routed upwards or
N O T E
downwards (depending on where the column is located). Route the capillaries directly
column (inlet) and waste assembly (outlet).
With the instrument accessory kit comes a 4-mm wrench and with the Sealing Kit a special
N O T E
adapter. Both together work as a torque wrench with pre-defined torque (maximum allowed
torque for the cell fittings is 0.7 Nm). It can be used to tight the capillary fittings at the flow
cell body. The wrench has to be plugged into the adapter as shown in
page 222.
Figure 76
222
Wrench plus Torque Adapter
Agilent 1260 Infinity DAD and MWD User Manual
Figure 76
on

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents