Agilent Technologies 1260 User Manual page 206

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

10
Maintenance
Maintenance of Standard, Semi-Micro or Micro Flow Cell
Next Steps:
9
Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
10
Reconnect the capillaries, see
11
Perform a leak test.
12
Insert the flow cell.
13
Replace the front cover
14
Perform a
"Wavelength Verification and Recalibration"
correct positioning of the flow cell.
206
"Exchanging a Flow Cell"
on page 183 or a
on page 200.
"Holmium Oxide Test"
Agilent 1260 Infinity DAD and MWD User Manual
on page 174 to check the