Agilent Technologies 1260 User Manual page 202

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

10
Maintenance
Exchanging a Flow Cell
N O T E
The label attached to the flow cell provides information
on part number, path length, volume and maximum
pressure.
If you want to replace flow cell parts,
Standard, Semi-Micro or Micro Flow Cell"
or
"Maintenance of High Pressure Flow Cell"
page 207.
7
Insert the flow cell capillaries into the union holder (top is
inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
Next Steps:
8
Perform a
"Wavelength Verification and Recalibration"
correct positioning of the flow cell.
9
Replace the front cover.
202
6
Insert the flow cell while pressing the flow cell holder.
"Maintenance of
on page 203
on
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
on page 183 or a
N O T E
"Holmium Oxide Test"
Agilent 1260 Infinity DAD and MWD User Manual