Agilent Technologies 1260 User Manual page 95

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

10 Enter the pump parameters mentioned under
11 Pump the water/acetonitrile (30/70 %) mobile phase through the column for
Agilent 1260 Infinity DAD and MWD User Manual
10 minutes for equilibration.
Table 22
• up to 8 signals (A to H) with individual wavelength
settings can be selected.
• spectrum settings,
"Spectrum Settings (DAD only)"
page 108
• stop and post time can be set (if required)
• depending on the application, the lamps can be
selected (one or both).
• peak width depends on the peaks in the chromatogram,
"Peakwidth Settings"
• autobalance to zero absorbance (on the analog output
plus offset) at begin and/or end of run.
• mechanical slit width can be changed for further
optimization,
"Slit Settings"
• margin for negative absorbance,
Absorbance Settings"
• Under More additional diagnostic signals can be added
for troubleshooting purpose, see section "Diagnostic
Signals" in the Service Manual.
• time table for programmable actions during the run.
NOTE: The Agilent G1315C/D and G1365C/D time
table can contain a maximum of 60 rows.
Using the Detector
Setting up an Analysis
on page 87.
on page 110
on page 112
"Margin for Negative
on page 113
5
on
95