Diagnosis Information on Agilent ChemStation

Diagnosis Information on Agilent ChemStation

The Agilent ChemStation provides certain information on various assemblies
in the detector.
1 Open the Diagnosis and change to detailed view.
2 Click on the assembly of interest and select "Update Variables Display".
Table 33
Table 33
Assembly
General
Vis lamp
UV lamp
Holmium Filter
Flow Cell
Micro Slit
186
Figure
on page 187 shows an example.
on page 186 lists the assemblies which provide detailed information.
Diagnosis - detailed information
Details
Product number, firmware revision, manufacturing date (of
intsrument or, if replaced, of main board), serial number,
accumulated on-time, spectrometer serial number.
Accumulated on-time, actual on-time, lamp switch on,
accumulated on time switches, switch on voltage/current, lamp
current, lamp voltage/current
Accumulated on-time, actual on-time, lamp ignitions,
accumulated on time reset, ignition voltage/current, lamp
voltage, lamp voltage/current.
Available with ID-tag lamp only.
product number, serial number, production date, last intensity
test
Filter movements, filter moves reset log
Available with ID-tag flow cells only.
Product number, production date, pathlength, max pressure,
serial number, volume, last cell test
Slit movements, filter moves reset log