Maintenance of High Pressure Flow Cell
1
Use a 4 mm hex key to unscrew the window assembly [1]
and remove the gasket [2] from the cell body.
If you want to replace the gasket only, continue with step 8,
"Maintenance of Standard, Semi-Micro or Micro Flow
Cell"
on page 203.
3
Follow the procedure
208
"Maintenance of Standard, Semi-Micro or Micro Flow Cell"
2
Use a tooth pick to remove the quartz window from the
window assembly.
on page 203 for reassembling.