Table 26
Peak Width (min)
<0.0025
>0.0025
>0.005
>0.01
>0.03
>0.05
>0.10
>0.20
>0.40
>0.85
Agilent 1260 Infinity DAD and MWD User Manual
Peak Width — Response Time — Data Rate
Response Time (sec)
0.025
0.05
0.1
0.2
0.5
1.0
2.0
4.0
8.0
16.0
Using the Detector
Special Settings of the Detector
Data Rate (Hz)
Detector
80
G1315C/G1365C only
80
G1315C/G1365C only
40
G1315C/G1365C only
20
G1315C/D and G1365C/D
10
G1315C/D and G1365C/D
5
G1315C/D and G1365C/D
2.5
G1315C/D and G1365C/D
1.25
G1315C/D and G1365C/D
0.62
G1315C/D and G1365C/D
0.31
G1315C/D and G1365C/D