Replacing Capillaries On A Semi-Micro And High Pressure Flow Cell - Agilent Technologies 1100 Series User Manual

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

8
Maintenance
Replacing Capillaries on a Semi-Micro and High Pressure Flow
Cell
When required
Tools required
Parts required
Preparations
1
Identify the inlet and outlet capillaries.
Inlet
capillary
136
If the capillary is blocked
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
For parts see
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
Outlet capillary
"Semi-Micro Flow Cell"
on page 158
"Exchanging a Flow Cell"
Notes
The fittings at the flow cell body are special types for
low dead volumes and not compatible with other
fittings.
When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with
a wrench). Otherwise damage of the flow cell body
or blockage may result.
1100 Series DAD and MWD User Manual
on page 121.