Agilent Technologies 1260 User Manual page 39

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Table 13
Performance Specifications G1315C/D and G1365C/D
Type
Specification
Flow cells
Standard: 13 µL volume, 10 mm cell path length
and 120 bar (1740 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell path length
and 120 bar (1740 psi) pressure maximum
Micro: 2 µL volume, 3 mm cell path length,
120 bar (1740 psi) pressure maximum
Semin-nano: 500 nL volume, 10 mm cell path
length and 50 bar (725 psi) pressure maximum
Nano: 80 nL volume, 10 mm cell path length and
50 bar (725 psi) pressure maximum
High pressure: 1.7 µL volume, 6 mm cell path
length and 400 bar (5800 psi) pressure
maximum
Prep SST: 3 mm cell path length and 120 bar
(1740 psi) pressure maximum
Prep Quartz:0.3 mm cell path length and 20 bar
(290 psi) pressure maximum
Prep Quartz: 0.06 mm cell path length and 20 bar
(290 psi) pressure maximum
Time programmable
Wavelength, polarity, peak width, lamp
bandwidth, autobalance, wavelength range,
threshold, spectra storage mode
Spectral tools
Data analysis software for spectra evaluation,
including spectral libraries and peak purity
functions
Control and data
Agilent ChemStation for LC (32-bit)
evaluation
Local Control
Agilent Instant Pilot (G4208A)
Analog outputs
Recorder/integrator: 100 mV or 1 V, output
range 0.001 – 2 AU, two outputs
Agilent 1260 Infinity DAD and MWD User Manual
Site Requirements and Specifications
Performance Specifications
Comments
See
"Optimization Overview"
All flow cells are equipped with RFID tags that
hold cell typical information.
For 1260 systems:
• Revision B.04.02 DSP2 or above
For 1100/1200 systems:
• Revision B.01.03 or above (G1315C/G1365C)
• Revision B.01.03 SR-2 / B.02.01 SR-2 or
above (G1315D/G1365D)
For 1260 systems:
• B.02.11 or above
For other systems:
• B.02.09 or above
2
on page 117
39