If another Agilent module is positioned on top of the
detector, route the tubing assembly waste from the
accessory kit behind the capillary holder and connect the
top end to the other module's waste outlet.
7
Remove the flow cell and establish a flow and observe for
leaks.
The installation of the detector is complete now.
Agilent 1260 Infinity DAD and MWD User Manual
Flow Connections to the Detector
6
Connect the capillary from the column to the capillary
holder (top). Connect the teflon waste tubing to the flow
cell outlet fitting (bottom) and the corrugated waste
tubing to the leak outlet.
8
Insert the flow cell, close the cover and replace the front
cover.