10
Maintenance
Nano Flow Cell - Replacing or Cleaning
5
The quartz body and the cell seal assembly can be
separated for cleaning purpose.
7
Replace the cell seal assembly onto the quartz body.
Always use a new seal assembly to exclude damage
during disassembling.
220
6
This figure shows the correct holding of the quartz body
and the cell seal assembly.
8
Slide the quartz body completely into the cell body to the
front stop (use for example a toothpick).
Agilent 1260 Infinity DAD and MWD User Manual