Replacing Capillaries on a Standard Flow Cell
1
Identify the inlet and outlet capillaries. To replace the inlet
capillary, continue with step "To replace the inlet
capillary, use a 4-mm wrench for the fitting."
3
To replace the inlet capillary, use a 4-mm wrench for the
fitting.
210
2
After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
4
Unscrew the cell body from the heat exchanger and the
heat exchanger from the clamp unit.
Agilent 1260 Infinity DAD and MWD User Manual