Bcd Cables - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

BCD Cables

One end of these cables provides a 15-pin BCD connector to be connected to
the Agilent modules. The other end depends on the instrument to be
connected to
Agilent Module to General Purpose
p/n G1351-81600
Agilent 1260 Infinity DAD and MWD User Manual
Wire Color
Pin Agilent
module
Green
1
Violet
2
Blue
3
Yellow
4
Black
5
Orange
6
Red
7
Brown
8
Gray
9
Gray/pink
10
Red/blue
11
White/green
12
Brown/green
13
not connected
14
not connected
15
Identifying Cables
BCD Cables
Signal Name
BCD Digit
BCD 5
20
BCD 7
80
BCD 6
40
BCD 4
10
BCD 0
1
BCD 3
8
BCD 2
4
BCD 1
2
Digital ground
Gray
BCD 11
800
BCD 10
400
BCD 9
200
BCD 8
100
+ 5 V
Low
12
261