Error Information
Detector Error Messages
Illegal Value From Air Inlet Temperature Sensor
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause
1
2
Wavelength Recalibration Lost
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause
1
2
156
The temperature sensor is defect.
Detector is exposed to illegal ambient
conditions.
The detector is new.
The detector has been repaired.
Suggested actions
Please contact your Agilent service
representative.
Verify that the ambient conditions are within the
allowed range.
Suggested actions
Recalibrate the detector.
Please contact your Agilent service
representative.