Agilent Technologies 1260 User Manual page 125

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Because the detector averages absorbance values that are calculated for each
wavelength, using a wide bandwidth does not negatively impact linearity.
The use of a reference wavelength is highly recommended to further reduce
baseline drift and wander induced by room temperature fluctuations or
refractive index changes during a gradient.
An example of the reduction of baseline drifts is shown in
page 125 for PTH-amino acids. Without a reference wavelength, the
chromatogram drifts downwards due to refractive index changes induced by
the gradient. This is almost completely eliminated by using a reference
wavelength. With this technique, PTH-amino acids can be quantified in the low
picomole range even in a gradient analysis.
Figure 55
Agilent 1260 Infinity DAD and MWD User Manual
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without
Reference
How to optimize the Detector
Figure 55
6
on
125

Hide quick links:

Advertisement

Table of Contents
loading