Maintenance
Exchanging a Flow Cell
Exchanging a Flow Cell
When
If an application needs a different type of flow cell or the flow cell needs repair.
Tools required
Two 1/4 inch wrenches for capillary connections
Parts required
#
1
1
1
1
1
Preparations
Turn the lamp(s) off.
Remove the front cover.
1
Press the release buttons and remove the front cover to
gain access to the flow cell area.
200
p/n
Description
G1315-60022
Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa)
G1315-60025
Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa)
G1315-60024
Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa)
G1315-60015
High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
Nano flow cell, refer to
"Nano Flow Cell - Replacing or Cleaning"