11
Parts for Maintenance

High Pressure Flow Cell

High Pressure Flow Cell

Item p/n
1
2
3
4
5
6
7
8
9
248
Description
G1315-60015
High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
Window assembly, comprises items 2, 3, 4, 5 and 6
79883-27101
Seal ring
1000-0953
Quartz window
79883-28802
Compression washer
5062-8553
Spring washers
79883-22404
Window screw
G1315-87325
Capillary IN (0.12 mm, 290 mm lg) including heat exchanger
G1315-87306
Capillary OUT (0.12 mm, 200 mm lg)
G1315-84901
Clamp unit
0515-1056
Screw M 2.5, 4 mm lg
for cell body/clamp
G1315-87312
Capillary column - detector 150 mm lg, 0.12 i.d.
G1315-87311
Capillary column – detector 380 mm lg, 0.17 i.d., (includes ferrule front
1/16", ferrule back 1/16" and fitting 1/16").
79883-68700
High pressure cell repair kit (includes 1 quartz window, 1 compression
washer, 5 spring washers, 2 seal rings)