Module Status Indicator - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

7
Troubleshooting and Diagnostics
Status Indicators

Module Status Indicator

The module status indicator indicates one of six possible module conditions:
• When the status indicator is OFF (and power switch light is on), the module
• A green status indicator, indicates the module is performing an analysis
• A yellow indicator indicates a not-ready condition. The module is in a
• An error condition is indicated when the status indicator is red. An error
• A red-blinking (modules with on-board LAN) or yellow-blinking (modules
• A fast red-blinking (modules with on-board LAN) or fast yellow-blinking
136
is in a prerun condition, and is ready to begin an analysis.
(run mode).
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, immediately after changing a set point), or while a
self-test procedure is running.
condition indicates the module has detected an internal problem which
affects correct operation of the module. Usually, an error condition requires
attention (e.g. leak, defective internal components). An error condition
always interrupts the analysis.
without on-board LAN) indicator indicates that the module is in resident
mode (e.g. during update of main firmware).
(modules without on-board LAN) indicator indicates that the module is in
boot loader mode (e.g. during update of main firmware). In such a case try
to re-boot the module or try a cold-start.
Agilent 1260 Infinity DAD and MWD User Manual