9
Test Functions

Test Chromatogram

Test Chromatogram

A pre-defined test chromatogram on the detector main board is processed
through ADC like normal signals from the diodes and can be used to check the
Analog-to-Digital Converter and the data handling system. The signal is
available at the analog output and on the LAN.
The run time of the chromatogram is depending on the setting for response time (peak
N O T E
width). If no stop time is set the chromatogram will repeat continuously.
Table 31
Response time
0.1 sec
0.2 sec
0.5 sec
1 sec
2 sec (default)
4 sec
8 sec
16 sec
Table 32
Peak
1
2
3
4
184
Test Chromatogram Settings (examples)
Set Run Time
0.28 min
0.6 min
1.2 min
2.3 min
4.5 min
9 min
18 min
36 min
Test Chromatogram - Peak Heights
Absorbance (approx.)
38 mAU
100 mAU
290 mAU
20 mAU
Agilent 1260 Infinity DAD and MWD User Manual