Replacing the Module's Firmware

When
The installation of newer firmware might be necessary
• if a newer version solves problems of older versions or
• to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
• to keep all systems on the same (validated) revision or
• if a new module with newer firmware is added to a system or
• if third part control software requires a special version.
Tools required
• LAN/RS-232 Firmware Update Tool or
• Agilent Diagnostic Software
• Instant Pilot G4208A (only if supported by module)
Parts required
#
1
Preparations
Read update documentation provided with the Firmware Update Tool.
To upgrade/downgrade the module's firmware carry out the following
steps:
1 Download the required module firmware, the latest LAN/RS-232 FW
2 To load the firmware into the module follow the instructions in the
Agilent 1260 Infinity DAD and MWD User Manual
Description
Firmware, tools and documentation from Agilent web site
Update Tool and the documentation from the Agilent web.
http://www.chem.agilent.com/scripts/cag_firmware.asp.
documentation.
Maintenance

Replacing the Module's Firmware

10
229