Shut-Down

An external instrument has generated a shut-down signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable cause
1
2
3

Remote Timeout

A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Probable cause
1
2
3
Agilent 1260 Infinity DAD and MWD User Manual
Leak detected in another module with a CAN
connection to the system.
Leak detected in an external instrument with
a remote connection to the system.
Shut-down in an external instrument with a
remote connection to the system.
Not-ready condition in one of the
instruments connected to the remote line.
Defective remote cable.
Defective components in the instrument
showing the not-ready condition.
Error Information
General Error Messages
Suggested actions
Fix the leak in the external instrument before
restarting the module.
Fix the leak in the external instrument before
restarting the module.
Check external instruments for a shut-down
condition.
Suggested actions
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the
instrument's documentation).