Test Failed
Dark-Current Test Evaluation
Probable cause
1
2
3
Agilent 1260 Infinity DAD and MWD User Manual
Defective slit assembly (stray light).
Defective detector main board.
Defective PDA/optical unit.
Dark-Current Test
Suggested actions
Run the
"Slit Test"
on page 166 (part of the
"Self-test"
on page 162).
Exchange the detector main board.
Exchange the optical unit.