Replacing Capillaries On A Semi-Micro, Micro And High Pressure Flow Cell - Agilent Technologies G1315B User Manual

Agilent 1200 series diode array and multiple wavelength detectors
Table of Contents

Advertisement

Replacing Capillaries on a Semi-Micro, Micro and High Pressure
Flow Cell
When required
Tools required
Parts required
Preparations
All descriptions in this procedure are based on the default orientation of the cell (as it is
N O T E
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not compatible
N O T E
with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.
1200 Series DAD and MWD User Manual
If the capillary is blocked
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
For parts
see"Semi-Micro Flow Cell"
page 150 or
"High Pressure Flow Cell"
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
on page 148,
"Micro Flow Cell"
on page 152.
"Exchanging a Flow Cell"
Maintenance
8
on
on page 103.
119