Technical Data
Technical data for DHHS components
In accordance with Federal Radiation Performance
Standards 21 CFR Subchapter J Paragraph 1020.30 (h)
X-Ray Controls and Generators
Nominal
line voltage:
Technique factors for maximum line current:
DCM operating mode, 125 kV
Generator ratings:
Fluoroscopy:
Pulsed Fluoroscopy:
Digital Radiography:
Digital Radiography (k=1):
(Single Shot Mode)
Road Map:
Subtraction:
Digital Cine Mode:
Siemens AG
Medical Solutions
Rated line
voltage:
100 V
110 V
120 V
127 V
200 V
230 V
240 V
tube voltage, 160 mA tube current.
p
SPR2-330.662.01
Rev. 02
Line-voltage-
regulation:
± 10%
± 10%
± 10%
± 10%
± 10%
± 10%
± 10%
125 kV
maximum voltage
p
15,2 mA maximum current
125 kV
maximum voltage
p
66,7 mA maximum current
125 kV
maximum voltage
p
14,3 mA maximum current
125 kV
maximum voltage
p
250 mA maximum current
125 kV
maximum voltage
p
15,2 mA maximum current
125 kV
maximum voltage
p
15,2 mA maximum current
125 kV
maximum voltage
p
250 mA maximum current
Page 1 of 2
05.06
CS PS SP
6
Maximum line
current:
11%
9%
8%
7%
7%
5%
5%
ARCADIS Avantic
6 - 1
6
6
28 A
26 A
24 A
22 A
14 A
12 A
12 A
6