Formulation

M
Purpose
With this application program you
can add weight values and
calculated values as components of
a formula to a totalizing memory.
You can use this application in
combination with any program
chosen from Application 1* (such as
counting, weighing in percent) and
one from Application 2 (check-
weighing, time-controlled functions) as
well as with the extra functions.
* = not with recalculation or 2
memory; cannot be activated
during formulation
Available Features
– Totalization of weight values and
calculated values
– Weigh in different components
toward zero to a total amount
defined by pressing the
Nom
key and entering the value through
the numeric keys
– Simultaneous storage of net and
calculated values, if available
– Optional configuration in the Setup
menu for loading weight values
and calculated values either from
Application 1 (e.g., counting,
weighing in percent) or from Appli-
cation 2 (checkweighing, time-con-
trolled functions)
– Totalizing memory for up to
65,535 values
– Transaction counter and current
total displayed in the text line
– Scale tared after a value
has been stored
– Manual input of the number of
individual weighing operations
and confirmation using the
soft key (target no. of operations
nDef). Result printed and memory
cleared after printout of nDef.
– Optional configuration in the Setup
menu to add the current weight,
with display accuracy, to the
current total by pressing the
key and generating a printout of
the result
– Optional configuration in the
Setup menu for stability-dependent
storage of the measured value:
Balance/scale functions, Stability
range
– Optional automatic storage of
measured values
Storage of measured value is
indicated by
you can place a load on the scale.
– Minimum load threshold for
automatic storage
– Press the
soft key to delete the
M-
last value added to the totalizing
nd
tare
memory. The transaction counter
value is reduced by one and
a printout is generated.
– Printout of an evaluation of results,
depending on the Application 1
or Application 2 parameters. Con-
figure the Setup menu to define the
information included on this print-
out.
– Press the key identified by
soft
(soft key) for a printout of an
intermediate evaluation after each
addition or a final evaluation
– A final evaluation is printed when
the formulation routine is ended
by pressing c, if no final
evaluation was generated by
pressing
MR
– Optional configuration in the Setup
menu to clear the totalizing
memory and reset the transaction
counter by pressing c or after an
evaluation is printed out
– Totalization data and transaction
counter data are stored in the
non-volatile memory
– Continue formulation after turning
the scale off and back on
Factory Settings of the Paramters
nDef
Automatic storage:
Minimum load for automatic storage:
10 digits
Source of data for auto storage:
Application 1
soft
M+
Operating the Scale
Evaluated values:
Evaluation mode,
Intermediate
evaluation, print
Printout of individual components:
Yes
Stability range:
Printout: Application-defined output:
;
indicates that
Uu
zz
Print on request then tare:
Soft Key Functions
M+
M-
MR
MR
nDef
Nom
Printout of Formulation Report
When an intermediate or final
evaluation is printed out, all results
up to this point are included.
Comp2 + 42.38 g
Tot.cp+184.89 g
Comp2:
Off
Tot.cp:
Net
key function:
MR
2 digits
Off
Add weight values or
application values to the
total in the totalizing
memory. The component
or transaction counter
value increases by one
each time you press this
key.
Delete the last value
added to memory. The
transaction counter value
decreases by 1. You
cannot delete previous
values by repeatedly
pressing this key.
Print or display an
intermediate or final
evaluation
Store the input number of
components
Press to enter target
component weight using
the numeric keys
Weight of the 2nd
component
Total of all components