Block Printout
You can have the results of a
calibration/adjustment procedure
printed out. You can configure
whether the printout is generated as
soon as the procedure is completed,
or whether a number of calibration/
adjustment procedures (up to 50)
are collected for a block printout.
Loading Stored Data:
Data for the block printout are stored
in battery-backed memory. The first
time you put the scale into operation, it
must remain connected to power for a
full day. These data remain in memory
for approx.3 months after the equip-
ment is disconnected from AC power.
Make sure to generate a printout befo-
re disconnecting the equipment for a
long period of time.
Block Printout of Calibration/
Adjustment Data
With the following Setup menu
configuration, you can store the data
from up to 50 calibration/adjustment
procedures and have them printed
on request:
FC Models:
– GLP/GMP calibration/adjustment
printout [1-17- ]
2 On request, from
record memory
FCA Models:
– Print GLP/GMP calibration/
adjustment record
On request, from data
memory
When the memory contains 50 data
records:
– additional records are output
automatically
If at least one block printout data
record has been configured, the
following soft keys are available after
you press the
soft key:
isoTST
The number of records is
Info
displayed in the text line
Print accumulated records
PrtPro
Delete accumulated
DelPro
records; records can only
be deleted after a printout
has been generated.
If a password has been
assigned in the Setup: Input
menu, you must enter either
the configured password
or the General Password
before you can delete the
records.
162
For internal calibration/adjustment,
the initialization mode of the proce-
dure is displayed in the Start line.
--------------------
13.05.1997
09:17
SARTORIUS
Mod.
FC6CCE-HX
Ser. no.
70419914
Ver. no.
01-35-06
ID
--------------------
24.04.1997
12:03
Start:
manual
Diff. +
0.001 g
External calibration
completed
25.04.1997
12:10
Start:
isoCAL/temp
Diff. +
0.001 g
Internal adjustment
completed
Diff. +
0.000 g
25.04.1997
18:30
Start:
Adj.time
Diff. +
0.001 g
Internal adjustment
completed
Diff. +
0.000 g
26.04.1997
Start:
manual
Diff. +
0.001 g
Internal adjustment
completed
Diff. +
0.000 g
27.04.1997
11:53
Start:
Ext.cal.
W ID
Nom.
+
500.000 g
Diff. +
0.001 g
External calibration
completed
Diff. +
0.000 g
--------------------
13.05.1997
09:17
Name:
--------------------
GLP header
List of Calibration/
Adjustment Procedures:
Example 1:
Internal calibration
Example 2:
isoCAL triggered by difference
in temperature
Example 3:
isoCAL at defined time
Example 4:
9:37
Internal calibration/adjustment
triggered manually
Example 5:
External calibration/adjustment
GLP footer