Net-Total Formulation/Second Tare Memory

With this application program you can weigh in components for formulation
of a mixture.
Preparation
Configure the Net-total formulation/Second tare memory application in the
operating menu: See "Configuring the Balance/Scale" Menu code 2.1. 3
Example
Step
1. Turn on the balance/scale
2. Place an empty container on
the balance/scale
3. Tare the balance/scale
4. Add the first component
5. Store the first component weight.
If the print format is set
to include data ID codes,
the following is printed
6. Add the next component
7. Store the 2nd component weight
8. Add further components,
if desired
9. Display total weight
and fill to desired final weight
Key
(or instruction)
e
t
v
v
As described
for Steps 5 and 6
c
Display
+ 65.0
g
+ 120,5
g
0.0
g
NET
N1
+
120.5 g
+ 70,5
g
0.0
g
NET
+ 191,0
g