Examples of Application Combinations

Example 1: Counting and checkweighing with statistical evaluation
You want to check a piece count, and have the results that lie within the tolerance range statistically evaluated and
printed as a ISO/GMP-compliant record.
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 3: Statistics: Automatic storage: On, first value at stability
Setup: App(lication parameters): Application 3: Statistics: Source of data for auto storage: Application 2
Setup: App(lication parameters): Application 3: Statistics: Evaluated value: Calculated
Setup: App(lication parameters): Application 3: Statistics: Evaluation mode, MR function: Intermediate evaluation, display+print
Setup: App(lication parameters): Basic application: Printout configuration: ISO/GLP/GMP printout: Always
Step
1. If necessary: turn on the scale
and enter the settings given above
2. Place reference sample quantity
on the scale
3. Initialize the scale
4. Remove reference sample quantity
Key (or instruction)
e
Place parts on the scale
soft key
Start
Unload the scale
Operating the Scale
Display/Output
149