Printout for Differential Weighing
Generating Configured Printouts
Automatically
The configured backweighing printout
is generated automatically after
backweighing, if one of the following
settings is selected under Setup:
Application 1: Differential weighing:
Generate printout:
Automatic after
backweighing
Auto after init.weigh +
backweigh
Auto after tare, init.
+backweigh.
Generating Configured Printouts
Manually
The configured individual printout is
generated when the p key is
pressed while there is a tare, initial or
backweight on the scale or when D
is pressed to toggle applications.
You can generate the configured
printout manually after backweighing
if you press the p key while the
display page for the results is shown.
To generate the configured statistics
printout, press the p key
– when the display page for statistics
is shown,
– when the samples with a desired
number of backweighing
operations is selected (for
example, statistics on all samples
with 2 backweighing operations)
The following printout is generated:
Configured Backweighing Printout (Example)
---------------------
16.11.1999
14:55:12
Lot
CH12345
Sample
ID
T1
+
23.45 g
N1
+ 125.57 g
R
(3)+ 103.68 g
R
+
82.57 %
D
-
21.88 g
D
-
17.43 %
Fact
+ 1.10345
D-Res
-
24.15 o
Ratio1
+
21.11 %
Ratio2
+ 121.11 %
---------------------
Operating the Scale
Dotted line
Date/Time
Lot ID
Sample number
14
Sample ID
CX88
Tare weighing (with PT1 selected)
Initial weight
Backweight (residue as weight)
Residue in percent
Loss as a weight
Loss in percent
Calculation factor
Calculated loss
Ratio 1
Ratio 2
Dotted line
103