Sartorius Factory Series Operating Instructions Manual page 102

Factory series electronic precision scales
Table of Contents

Advertisement

Differential Weighing:
Defining the Weighing Sequence
You can choose from among four
sequences for measuring tare weights,
initial sample weights and back-
weighed residue ("backweight")
during differential weighing:
1. Individual Weighing
Tare weight, initial weight and
backweight are measured in that
order.
Measured weights/values
Tare
Init.wt.
1
1
2. Consecutive Individual Weighing
Several individual weighing
routines (see above) are performed
in series.
Measured weights/values
Tare
Init.wt.
1
1
Tare
Init.wt.
2
2
Tare
Init.wt.
3
3
102
3. Combined Weighing
The tare and initial weight, in that
order, of each sample is measured
first, then the backweight of each
sample is measured.
Measured weights/values
Tare
1
Tare
2
Backwt.
1
Tare
3
4. Serial Weighing
First the tare weight for each
sample is measured, then the initial
weight of each sample is measured
in the same order that their tare
weights were measured, and then
all backweights are measured.
Backwt.
Measured weights/values
1
Tare
Backwt.
1
2
Tare
2
Backwt.
3
Tare
3
You can define the weighing sequen-
ce in the Setup menu or by pressing
the
(if the "Weighing
Wg.seq
sequence key" option is activated).
Factory Settings
Weighing sequence:
Group weighing
Tare weighing:
Result with decimal point:
2 decimal places
Autosave values:
Minimum load for autosave:
Init.wt.
Backwt.
1
1
10 digits
Save statistics:
Init.wt.
Backwt.
Generate printout:
2
2
Automatic after
backweighing
Include sample ID in the text line:
Init.wt.
Backwt.
No
3
3
Wg. seq. key:
Yes
Clear sample after individual weight,
result + unload:
No
Last residual weight saved as the initi-
al weight:
Init.wt.
Backwt.
1
1
Init.wt.
Backwt.
2
2
Init.wt.
Backwt.
3
3
Yes
No
No
No