Preparation
To calculate a piece count, the average weight of one piece must be known.
This average piece weight can be entered into the Counting program in one of
three ways:
– Enter the average piece weight using the numeric keys and store it;
– The last reference sample quantity entered is loaded and displayed when
you turn on the scale. Place the same number of parts on the scale and
initialize the Counting program;
– When the automatic initialization parameter (see previous page) is on
(Setup: Printout: Application-defined output: Auto print upon initialization:
All values), the scale goes into the "Counting" mode when you turn it on and
loads the last average piece weight and corresponding reference sample
quantity that were entered or calculated.
Reference Sample Updating
You can have the average piece weight updated during counting (with
the piece count displayed) if "AWP update" is set to "manual" or "automatic"
in the Setup menu. Manual updating can only be performed when
the
soft key is displayed. Reference sample updating must be
Update
completed before using an application program from Application 3.
The
soft key is displayed when:
Update
– the scale has reached stability (stability symbol displayed)
– the current piece count is less than double the original piece count
– the current piece count is less than 100
– the internally calculated piece count (e.g., 17.24) differs from the nearest
whole number (here: 17) by less than 0.3
Reference updating can be repeated several times with an approximately
doubled piece count.
§ To perform reference updating: Press the
§ Turn on the scale: Press e
> Sartorius logo is displayed, self-test is performed
§ Select the "Counting" program in the Setup menu: Press M
FC Models:
§ Select the Application menu: Press the
§ Select
Application 1
FCA Models:
§ Select
Application parameters
then the O soft key once
§ Select
Application 1 (basic settings)
§ Select
: Q or q soft key, repeatedly
Counting
§ Confirm
: O soft key
Counting
Counting
Average – average
piece weight calculation
Average piece
weight updating
o = factory setting
see also the "Application Menu (Overview)" in the chapter entitled
"Configuring the Scale"
§ Save settings and exit the Setup menu: Press the oo soft key
soft key
Update
soft key
App
: O soft key
: press the q soft key 2 x,
: press the O soft key
o Display accuracy
+1 decimal place
+2 decimal places
Off
Manual
o Automatic
Operating the Scale: FCA
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during
alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
§ Press the
soft key
isoTST
> See "Calibration/Adjustment"
for further instructions
Toggling to the Next Application
§ Press D
> See the section on the
corresponding application
program for further instructions
Setup (setting parameters)
§ Press M
> See "Configuring the Scale"
for further instructions
Turning Off the Scale
§ Press e
> The scale shuts off
> The display goes blank