Step
The scale delays starting the
subweighing operation until
three successive subweights lie
within the range defined for
a "calm" animal
After 20 subweighing operations
(
: number of current subweigh
n
: arithm. average, net value)
x-Net
7. Store result and activate autom.
storage by pressing the M+ soft key
(automatic storage is not
active here*)
8. Unload the scale
9. Weigh all 7 animals
The next weighing operation
starts automatically; the result
is stored automatically in the
Statistics program
10. View display, then print
*
The first time you store a value
after the Statistics memory
has been cleared, storage
must be initiated manually,
by pressing the
soft key.
M+
The subsequent values in
the statistics series will be
stored automatically.
152
Key (or instruction)
When this criterion is met, the
subweighing series begins
soft key
M+
Remove animal from cage
Place one animal
after another in the cage
D
soft key
MR
p
Display/Output
n
x-Net +
31.70 g
1