Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of –5 g and +10 g. Printout of upper and lower
tolerance limits. Weighed values are printed out automatically when stability is reached and the weight value is within
the checkweighing range.
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 2: Checkweighing: Automatic printout of OK values: On
Step
1. Turn on the scale and configure
the settings as indicated above
2. Delete previous values, if necessary
3. Prepare a container
for the samples
4. Tare the scale
5. Enter initialization values
6. Accept target value
(in this example: 170 g)
7. Save target and unload
the scale
8. Enter value for lower limit
(170 g – 5 g) and save
116
Press key(s)
(or follow instructions)
e
c
Place empty container
on the scale
t
soft key
Param.
Place ideal sample
in container
l soft key
Remove ideal sample
from the scale
1 6 5
l soft key
Display/Output