Practical Example
Document the evaporated amount of a sample with defined surface, temperature and air pressure at preset intervals
of 1 minute, 30 seconds.
Settings (changes in the factory settings required for this example):
Setup: App(lication parameters): Application 2: Time-controlled functions
Setup: Balance/scale functions: Taring: Without stability
Setup: Printout: Application-defined output: Stability parameter: Without stability
Step
1. Turn on the scale and configure
the settings as indicated above
2. Delete stored values, if necessary
3. Place container with sample
on the scale and tare
4. Enter time interval: 1 minute,
30 seconds
5. Store time interval
6. Begin documentation
(Time remaining until the
next printout is displayed
in the text line)
Printout of evaporated amount
every 1
1
/
minutes
2
7. Stop the documentation procedure
120
Press key(s)
(or follow instructions)
e
c
t
1 . 3 0
soft key
Interv
soft key
Start
soft key
Stop
Display/Output
6
isoTST
6
6
isoTST
6
isoTST