Sartorius Factory Series Operating Instructions Manual page 66

Factory series electronic precision scales
Table of Contents

Advertisement

Printing an ISO/GLP/GMP-
compliant Record
In the Setup menu, you can configure
whether
– no ISO/GLP/GMP-compliant
record is printed
(
,
)
Off
[8-10-1]
– an ISO/GLP/GMP-compliant
record is printed only after
calibration/adjustment
(
Only for
calibration/
adjustment, [8-10-2]
– every printout is an
ISO/GLP/GMP-compliant record
(
,
Always on
[8-10-3]
Undoing All Parameter Changes –
Reset Function
There is a factory setting for each
parameter. In the Setup menu, you
can configure whether
– factory settings are restored after
exiting Setup (
Factory
settings, [9-1-1]
66
)
)
)