LabMaster Channel Setup

You can adjust the sensitivity and the positioning of the waveforms and set
the color of the trace for up to 80 channels on the Channel Setup dialog.
The example shown below illustrates Channel Setup with a 20 channel
LabMaster.
1.
Choose Vertical → Channel Setup... from the menu bar to open the
Channel Setup dialog.
2.
Touch inside the checkbox next to a channel to display the trace.
3.
To adjust sensitivity and position, turn the V
knob for the selected channel.
OR
Touch inside the Volts/Div field next to the channel and enter a
value. Click the keypad button to enter a value using the pop-up
keypad, or use the up/down arrows.
NOTE
: The set voltage is shown on the trace descriptor label and in the
Volts/Div
4.
To set the offset, turn the V
directly above the channel button whose waveform you want to
move vertically. Or, you can touch inside the Offset field next to the
channel and type in a value on the pop-up keypad. To set the vertical
offset to zero, touch the Zero Offset button directly below the Offset
field.
5.
To change the color assigned to a channel, click on the color box and
choose a color.
NOTE
: Front panel knobs can be used to adjust
for channels 5-80, provided the trace descriptor label is active for the
corresponding channel.