Running Pass/Fail Test
Initial Setup
1.
Touch Analysis → Pass/Fail Setup... on the menu bar.
2.
Touch the Actions tab.
3.
Touch the Enable Actions checkbox. This causes selected actions to
occur on your waveform's passing or failing a test.
4.
Touch the Summary View checkbox to enable a line of text (as
follows) showing a concise status of the last waveform and keeping a
running count of how many sweeps have passed.
5.
Touch inside the Pass If control, and select a Boolean condition from
the choices shown.
6.
If you want to set up the test to end after a finite number of sweeps,
touch the Stop Test checkbox. Then touch inside the After data entry
control and provide a value using your preferred input control
method. Additional information on using the touch screen controls
can be found in the Dialog Area.
7.
Use the Pass or Fail button on the If section of the dialog to set
actions taking place when your waveform passes or fails the test.
8.
The Then section of the dialog provides stop test, sound alarm, print
result, emit pulse, or save (the waveform) actions. If you want to
have the results printed (and your oscilloscope is not equipped with a
printer) make sure your instrument is connected to a local or
network printer.
9.
If you want to save your waveform automatically, touch the Save
Setup... checkbox. This takes you out of the current dialog and opens
the Save Waveform dialog.
10.
Test Pass/Fail conditions by touching the Force Actions Once button.
Press the Clear All button to quickly uncheck all checkboxes and
make new selections.
922561-00 Rev A
See Print Setup Dialog in the online Help.
Getting Started Manual
127