You can use Sequence mode in remote operation to take full
advantage of the instrument's high data-transfer capability.
In Sequence mode, the complete waveform consists of a number of fixed-
size segments acquired in single-shot mode (see the instrument
specifications for the limits). The oscilloscope uses the sequence timebase
setting to determine the capture duration of each segment as 10 x time/div.
With this setting, the oscilloscope uses the desired number of segments,
maximum segment length, and total available memory to determine the
actual number of samples or segments, and time or points.
60
How the instrument captures segments
922561-00 Rev A