Teledyne LabMaster 10 Zi Series Getting Started Manual page 133

Hide thumbs Also See for LabMaster 10 Zi Series:
Table of Contents

Advertisement

1.
Touch Analysis → Pass/Fail Setup... from the menu bar.
2.
Touch one of the Qx tabs and a setup dialog for the position is
shown.
3.
Touch Condition and select DualParamCompare.
4.
Touch inside the Source1 and Source2 fields and select a source from
the pop-up menu.
5.
Touch inside the ParamCompare mini-dialog field and select a source
from the pop-up menu.
6.
Touch inside the Compare Values field and select All or Any from the
pop-up menu.
By selecting All, the test are true only if every waveform falls within
the set limit. Selecting Any makes the test true if just one waveform
falls within the limit.
7.
Touch inside the Condition field on the ParamCompare part of the
dialog and select a math operator from the pop-up menu.
8.
Touch inside the Limit field and enter a value, using the pop-up
numeric keypad. This value takes the dimension of the parameter
that you are testing. For example, if you are testing a time parameter,
the unit is seconds.
9.
If you choose either
from the Condition menu respectively, touch inside the Delta field
and enter a value.
Also see Mask Testing (on page 128).
922561-00 Rev A
WithinDeltaPct or
Getting Started Manual
WithinDeltaAbs
125