Teledyne LabMaster 10 Zi Series Getting Started Manual page 84

Hide thumbs Also See for LabMaster 10 Zi Series:
Table of Contents

Advertisement

LabMaster 10 Zi Oscilloscopes
If you perform ChannelSync Cal on multiple gain settings, perform it
first on the 100 mV/div setting, then perform it on other gain
settings (with "Reference Correction" values entered as appropriate
for those other gain settings.)
For the highest possible accuracy across the full input range of the
oscilloscope, a ChannelSync Cal should be performed on at least one
gain setting in each of the following gain ranges:
2.92mm up-to-36 GHz Inputs
10.1 to 11.8 mV/div
11.9 to 17.0 mV/div
17.1 to 20.5 mV/div
20.6 to 30.0 mV/div
30.1 to 39.5 mV/div
39.6 to 50.0 mV/div
50.1 to 75.0 mV/div
75.1 to 90.0 mV/div
90.1 to 100.0 mV/div
100.1 to 126.0 mV/div
126.1 to 158.0 mV/div
158.1 to 205.0 mV/div
205.1 to 250.0 mV/div
250.1 to 315.0 mV/div
315.1 to 500 mV/div
If you add cables or probes to your setup subsequent to running the
ChannelSync Cal, you must manually deskew these items in the
Channel dialog to account for their propagation delays. The values
you provide on the Channel dialog (in the deskew section) are
additional to values shown on the ChannelSync Calibration dialog.
76
≤10 mV/div
1.85mm up-to-65 GHz Inputs
≤10 mV/div
10.1 to 11.8 mV/div
11.9 to 14.2 mV/div
14.3 to 17.0 mV/div
17.1 to 20.5 mV/div
20.6 to 25.0 mV/div
25.1 to 30.0 mV/div
30.1 to 39.5 mV/div
39.6 to 50.0 mV/div
50.1 to 63.0 mV/div
63.1 to 80 mV/div
922561-00 Rev A