4.
Touch the Enable Timeout checkbox.
5.
Touch the Timeout data entry control and provide a timeout value.
NOTE
: Use the sequence mode timeout to automatically interrupt the
sequence acquisition if the timeout value is exceeded without a valid trigger.
The timeout period accounts for instances when a
miscount occurs for some reason and the oscilloscope waits indefinitely for
an unforthcoming segment. During that time, no oscilloscope functions are
accessible. By means of a timeout value, however, the acquisition will be
completed, the waveform displayed, and control of the oscilloscope
returned to the user after the timeout has elapsed.
6.
Under Display Settings, touch the Display mode control, and select a
sequence mode display from the pop-up menu.
NOTE
: Once a single acquisition has started, you can interrupt the
acquisition at any time by pressing the
time or by pressing the
already acquired will be retained in memory.
7.
Touch the
922561-00 Rev A
S
front panel button. In this case, the segments
TOP
S
trigger front panel button.
INGLE
Getting Started Manual
Num Segments
S
front panel button a second
INGLE
63