Getting Started Manual
Quick Access to Pass/Fail Setup Dialogs
You can quickly gain access to the 1) Pass/Fail dialog, 2) Qx setup dialog, or
the 3) Actions dialog by touching one of the sections below the grid display.

Removing a Mask from the Display

Quickly remove a mask by accessing the Qx dialog. On the Load Mask right-
hand dialog, click the Delete button.
Mask testing can be done using an existing mask, or by using a mask created
from your actual waveform, with vertical and horizontal tolerances that you
define. Existing masks can be loaded from a floppy disk or from a network.
You can set your mask test to be True for waveforms All In, All Out, Any In,
or Any Out. For example, if you select All In, the test becomes False if even a
single waveform falls outside the mask.
Masks that you create from your waveform can be confined to just a portion
of the trace by use of a measure gate. Refer to the Measure Gate topic in
the online Helptopic for more information on how this feature works.