Repeat these steps for each Acquisition Module in your system.
PLEASE NOTE THE FOLLOWING:
PCIe 4 Lane Cable plugs are keyed with a single groove along one
wide side of the plug. The plug must be inserted into the socket with
the groove aligned properly.
Be sure to connect your Acquisition Modules to the correct channel
groupings on the Master Control Module; First Acquisition Module is
connected into DATALINK CHANNEL 1-4, second into the DATALINK
CHANNEL 5-9 input,
third into the DATALINK CHANNEL 9-12 input,
fourth into the DATALINK CHANNEL 13-16 input,
and last into the DATALINK CHANNEL 17-20 input.
If you are connecting less than 5 of the Acquisition Modules, the
PCIe 4 Lane Data Inputs on the back of the Master Control Module
may not be skipped and must be connected in consecutive order
from the PCIe 4 Lane Data Outputs on the back of the corresponding
slaves.
Pull gently on the green pull tab on the cable plug for removal.