Cursors; Quickly Displaying Cursors - Teledyne LabMaster 10 Zi Series Getting Started Manual

Hide thumbs Also See for LabMaster 10 Zi Series:
Table of Contents

Advertisement

LabMaster 10 Zi Oscilloscopes

Cursors

Cursors are important tools that aid you in measuring signal values. Cursors
are markers - lines, cross-hairs, or arrows - that you can move around the
grid or the waveform itself. Use cursors to make fast, accurate
measurements and eliminate guesswork. There are two basic types:
Horiz(ontal) (generally Time or Frequency) cursors are markers that
you move horizontally along the waveform. Place them at a desired
location along the time axis to read the signal's amplitude at the
selected time.
Vert(ical) (Voltage) cursors are lines that you move vertically on the
grid to measure the amplitude of a signal.
Each cursor locates a point on the waveform. The cursor values can be read
in the descriptor label for the trace. Use the Position data entry controls at
the right side of the dialog to place the cursors precisely. The Readout icons
depict what is being measured for each mode.

Quickly Displaying Cursors

Quickly turn on/off cursors from either the touchscreen or the Front Panel:
On the display, touch Cursors on the menu bar and make a selection.
Off, Horizontal Abs, Horizontal Rel, Vertical Abs, Vertical Rel, and
BothRel options are available on the menu.
On the Front Panel, press the Cursor Type button repeatedly to scroll
through all the cursor types available on your instrument. With the
desired cursor type selected, turn the Cursors knobs to adjust the
position of the cursor.
PLEASE NOTE THE FOLLOWING:
The bottom cursor knob adjusts relative cursors. It does not work
with absolute cursor types.
Previously setup cursor positions are preserved for subsequent
displays. Changing cursor positions or measurement modes are
always available using touch screen or front panel control methods as
previously mentioned.
104
922561-00 Rev A

Advertisement

Table of Contents
loading

Table of Contents